Mistake #9: Drainage Error
 Description of the problem
 Clogged drain.
 What can I do to solve the problem?
-  Check the wastewater level and pipe connections.
 - Check the level of used water in the bottle with the red cap. If the bottle is full, discard the used water.
 -  Check that the red pipe is not kinked and that it is pointing downwards along its entire length.
 -  Check that there is no debris in the outlet opening (inside the chamber).
 - Click the button below to contact us.
 